Protection Through Detection
MeshGuard
MeshGuard
• MeshGuard Zero Calibration
– Before calibration, make
sure the ambient air is
clean from the target gas
and check that you have
the right span gas.
– Enter the program menu by
pressing [M] and [N/-] at
the same time.
– First screen is zero
calibration. Press [Y/+] to
zero calibrate or [N/-] to go
to the next screen.